RecallDepth

Boston Scientific IQ Guide Wire. Catalog No. 38950-01J0, 185 cm, J-Tip, Single Pack. Sterile EO. Manufactured for: Boston Scientific Corporation, One Boston Scientific Place, Natick, MN 01760 USA. Boston Scientific Guide Wires are steerable guide wires available in a variety of lengths and diameters. The distal tip is shapeable, or, alternatively, is available in a preshaped ''J'' tip.

This recall has been terminated (originally issued July 12, 2006).

Company
Boston Scientific
Recall Initiated
May 30, 2006
Posted
July 12, 2006
Terminated
June 28, 2007
Recall Number
Z-1214-06
Quantity
1455 units total
Firm Location
Maple Grove, MN
Official Source
View on FDA website ↗

Reason for Recall

Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.

Distribution

Nationwide (AR, AZ, FL, GA, IA, ID, IL, IN, KY, LA, MA, MI, MN, MO, NC, NH, NJ, NM, NV, OH, OK, PA, TN, TX, UT, VA, WI, WV) and Puerto Rico. Chile, Germany, Taiwan.

Lot / Code Info

Catalog No. 38950-01J0, Box Lot No. AND Inner-Pouch Lot/Batch #: 8435441/8416147, 8445221/8395559, 8456936/8422573.

Root Cause

Other

Action Taken

On May 30, 2006, Boston Scientific notified customers of the IQ Guide Wire manufactured from February 3, 2006, through March 15, 2006. BSC requested a reply whether or not they still have product in their institution. Replacements will be issued for returned product. UPDATE: A second recall notification was sent to affected customers 07/07/06. This second notification included Inner Pouch Number (UPN) and Inner Pouch Lot/Batch #.