Programming software (all versions) for Polymedco PolyTiter Immunofluorescent Titration System and used with DiaSorin ANAFLUOR Indirect Fluorescent Antibody Test.
- Company
- Diasorin Inc.
- Recall Initiated
- September 24, 2004
- Posted
- May 10, 2005
- Terminated
- July 22, 2006
- Recall Number
- Z-0807-05
- Quantity
- 13
- Firm Location
- Stillwater, MN
Reason for Recall
A problem with software for the PolyTiter Immunofluorescent Titration System in that no flags or warnings appear when PolyTiter Calibrator Values are re-entered or changed (including values changed to unacceptable values) after the initial curve has been generated, if the 'refresh button' has not been pressed. The graph may appear valid; however erroneous results may occur.
Distribution
OH, NJ, KY, TX, NC, FL, VA, WA, AL, CO, MO, AZ , CA , CT & NV
Lot / Code Info
all versions of software
Root Cause
Other
Action Taken
Customers were sent a Technical Bulletin describing the correct procedure to be used when entering data into the Poly/Titer tiration system, until the updated software is installed