Software written for a specific automated microtiter plate instrument, Bio-Tek ELx800 reader for use with the ENA 6 Screen ELISA kit assay.

This recall has been terminated (originally issued April 6, 2005).

Company
Diasorin Inc.
Recall Initiated
April 20, 2004
Posted
April 6, 2005
Terminated
July 22, 2006
Recall Number
Z-0672-05
Quantity
29
Firm Location
Stillwater, MN
Official Source
View on FDA website ↗

Reason for Recall

An error in the Bio-Tek ELx800 microplate reader programming specifications for the ENA screen Assay exists in that an equivocal (borderline) zone is missing.

Distribution

CT, FL, NY, WA, TX, AR, TN, CANADA

Lot / Code Info

software versions 9 thru 12

Root Cause

Other

Action Taken

Customers were sent a notification letter informing customers of an error in the Bio-Tek ELx800 programming specifications. Customers were asked to repeat testing for historical runs for RSLT (S/CO) that read from 0.95 to 1.0. A response form is asked to be faxed back to DiaSorin.