Software written for a specific automated microtiter plate instrument, Bio-Tek ELx800 reader for use with the ENA 6 Screen ELISA kit assay.

Company
Diasorin Inc.
Recall Initiated
April 20, 2004
Posted
April 6, 2005
Terminated
July 22, 2006
Recall Number
Z-0672-05
Quantity
29
Firm Location
Stillwater, MN

Reason for Recall

An error in the Bio-Tek ELx800 microplate reader programming specifications for the ENA screen Assay exists in that an equivocal (borderline) zone is missing.

Distribution

CT, FL, NY, WA, TX, AR, TN, CANADA

Lot / Code Info

software versions 9 thru 12

Root Cause

Other

Action Taken

Customers were sent a notification letter informing customers of an error in the Bio-Tek ELx800 programming specifications. Customers were asked to repeat testing for historical runs for RSLT (S/CO) that read from 0.95 to 1.0. A response form is asked to be faxed back to DiaSorin.