ARCHITECT Processing Module, manufactured by Abbott Laboratories, Irving, Texas

This recall has been terminated (originally issued July 20, 2004).

Company
Abbott Laboratories, Inc
Recall Initiated
May 8, 2002
Posted
July 20, 2004
Terminated
November 26, 2003
Recall Number
Z-0172-04
Quantity
1,167 units worldwide
Firm Location
Irving, TX
Official Source
View on FDA website ↗

Reason for Recall

If the interlock cable connected to the door lid sensor is disconnected or fails, the lid is detected as closed even though it may be open.

Distribution

States of Virginia, Illinois, New York, Texas, Oklahoma, Pennsylvania, ILlinois, Michigan, California, New Hampshire, Ohio, Florida, Maryland, Nevada, Georgia, Minnesota, Colorado, Arizona, Missouri, Louisiana, Oregon, Wisconsin, Hawaii, Connecticut, New Hampshire, Massachusetts, Washington, South Carolina, Mississippi, Alabama, and Tennessee, and the following foreign countries: Austria, Denmark, Germany, Italy, Poland, South Africa, Sweden, England, Colombia, Australia, Japan, New Zealand, Belgium, Finland, Greece, Netherlands, Portugal, Romania, Turkey, Canada, Mexico, Hong Kong, Korea, Taiwan, Czech Republic, France, Israel, Norway, Saudi Arabia, Spain, United Arab Emirates, Brazil, Venezuela, India, Malaysia, and Thailand.

Lot / Code Info

All Serial numbers

Root Cause

Other

Action Taken

Firm sent technical service bulletins with corrective instructions to consignees with all existing ARCHITECT i 2000 processing modules in February 2003.