Abbott FreeStyle Blood Glucose Monitoring System Test Strips for use with FreeStyle Blood Glucose Monitoring Meter

Company
Abbott Diabetes Care, Inc.
Recall Initiated
April 24, 2008
Posted
September 11, 2008
Terminated
March 29, 2011
Recall Number
Z-1813-2008
Quantity
2954900 strips
Firm Location
Alameda, CA

Reason for Recall

Flaws in strip voltage continuity may result in frequency of an Error 3 ("Er3") message display and unreadable strips. This issue has the potential for creating a delay in generating blood glucose results and to delay diabetes management, potentially leading to either hypo or hyperglycemia and their associated complications.

Distribution

Class II Recall - Worldwide Distribution --- including USA and countries of Sweden, Norway, Netherlands, Luxembourg, Switzerland, Israel and Denmark.

Lot / Code Info

Part numbers/Serial numbers/Exp. Date: CAT12050-01/0618625/Jul08; M70449-03/0703325/Feb09; CAT12450-01 (70345-02)/0618129/Jun08; CAT70450-01/0700815/Jan09; CAT70447-02/0628646/Oct08; CAT12450-01 (70345-02)/0627030/Sep08; CAT70450-02/0703019/Jan09; CAT12050-02/0703118/Jan09; CAT12050-03/0715129/May09

Root Cause

Process control

Action Taken

Recall initiated on April 24, 2008. An Urgent Medical Device Correction was issued to customers. Consignees were notified, along with registered users of the FreeStyle Freedom meters, via recall letters, visits or phone calls. Distributors will be notified, along with the creation of customer lists and tracking of communications to sub-recall. If you have any questions call Tina Mazurkiewicz at 1-800-777-6565.