RecallDepth

Ascent POCTS (Posterior Occipital Cervical Thoracic System) Part Numbers: 65-33 XX Series (65-3310,-3312, -3314, -3318, -3320-, -3322, -3324, -3326, -3328, and -3330), Multi-Axial Screw, 4.0mm x 26mm;Implant grade Titanium Alloy; Blacksone Medical Inc., 90 Brookdale Drive, Springfield, MA 01104.

This recall has been terminated (originally issued December 11, 2007).

Company
Blackstone Medical, Inc.
Recall Initiated
May 15, 2007
Posted
December 11, 2007
Terminated
June 6, 2008
Recall Number
Z-0363-2008
Quantity
319 screws ( 212 domestic, 107 international)
Firm Location
Wayne, NJ
Official Source
View on FDA website ↗

Reason for Recall

Manufacturing Defect: The saddle of the screw's body relative to the keyway exceeds the tolerance outlined in the device specifications. The extent to which the tolerance is exceeded varies from part to part, in the extreme the tolerance is exceeded by (+) 50%. As a consequence, there is a potential for set-screw cross-threading during implantation. Cross-threading could in turn lead to a comprom

Distribution

Worldwide: USA and countries of China, Cyprus, Turkey and Switzerland.

Lot / Code Info

Part Numbers: 65-3310,-3312, -3314, -3318, -3320-, -3322, -3324, -3326, -3328, and -3330; Lot numbers: 009, 010, 011, 018, 019, 031, 048, 064, 084 and 107.

Root Cause

Nonconforming Material/Component

Action Taken

Blackstone Medical, Inc., field representatives were notified via phone, e-mail or fax on 4/30/07. Attention: Notification of Product Removal Letters were sent on May 15, 2007, to distributors and implanting surgeons requesting return of non-implanted stock.

More recalls by Blackstone Medical, Inc.