ABL800 FLEX Series analyzer with FLEXQ module-equipped with software version 5.2 to 5.27

This recall has been terminated (originally issued April 6, 2007).

Company
Radiometer America Inc
Recall Initiated
January 16, 2007
Posted
April 6, 2007
Terminated
May 8, 2012
Recall Number
Z-0705-2007
Quantity
64 in U.S. & 1,281 to foreign accounts
Firm Location
Westlake, OH
Official Source
View on FDA website ↗

Reason for Recall

In some situations pre-registered patient data is not reset when leaving the pre-registration mode in the software. This occurs when a Safe Pico blood sampler associated with a specific patient is pre-registered in the analyzer by scanning a barcode.

Distribution

Worldwide Distribution-USA including states of NY, KS, SD, OH, TX, KY, NC, IL, MO, AL, UT, CO, CA, NM, and MI and Country of Denmark

Lot / Code Info

There are no specific lot codes. (devices equipped with software version 5.2 to 5.27)

Root Cause

Other

Action Taken

Radiometer America issued a recall letter to their U.S. customers dated 1/16/2007. The letter informs the customers of this software anomaly and the conditions underwhich analyzer results can be stored with incorrect Patient Demographics. The letter goes on to inform the customers that a Radiometer Service Representative will be visiting each customer location shortly in order to upgrade the affected software. In the meantime, the letter recommends that the customer(s) not pre-register their samples. The letter is accompanied by an attached FAX FORM which the customer is asked to sign and complete in order to verify receipt of the recall letter and to fax back to the recalling firm.