OEC 9800 Plus Fluoroscopic X-ray System,

This recall has been terminated (originally issued March 22, 2007).

Company
Ge Oec Medical Systems, Inc
Recall Initiated
February 9, 2007
Posted
March 22, 2007
Terminated
June 26, 2007
Recall Number
Z-0558-2007
Quantity
11 units of all models
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

During routine service, a cover may have been installed without the required proper lead shielding.

Distribution

Nationwide to AL, AZ, IA, OH, TX, VA, and WV.

Lot / Code Info

Serial numbers: 82-7092-MH, 89-0247-RC, 89-0599, 89-0972, 89-1001, 89-1685, 89-2383, 89-2877, 89-2907, 89-2987, 89-3330.

Root Cause

Other

Action Taken

Consignees were notified by telephone on 02/09/2007 and told to discontinue use of the units until they could be inspected by GE personnel. GE personnel visited all 8 sites on 02/12/2007.