Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763.

Company
Terumo Cardiovascular Systems Corp
Recall Initiated
November 1, 2005
Posted
June 30, 2007
Terminated
June 23, 2009
Recall Number
Z-0598-2007
Quantity
485 of both model numbers
Firm Location
Ann Arbor, MI

Reason for Recall

The power manager subsystem may not initialize properly during start-up, which will cause the battery status monitoring and battery switching operations to not be fully functional.

Distribution

Worldwide distribution ---- Nationwide and countries of Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Philippines, Republic of Georgia, Russia and Saudi Arabia.

Lot / Code Info

Serial numbers 0011 through 0363 may be affected. Serial numbers 0011, 0014, 0023, 0038, 0043, 0047, 0119, 0145, 0245, 0253, 0254, 0256, 0260, 0265, 0278, 0280, 0284, 0295, 0305, 0310, 0317, 0333, 0341, 0343, 0351, 0356 and 0360 have been corrected and the other serial numbers listed will be corrected.

Root Cause

Other

Action Taken

Firm repaired the affected units beginning in November 2005, upon receipt of complaints of failure. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that they will be visited and their units upgraded as replacement components become available.