D711, Reference Membranes, Part #942-058, used with the ABL835 Blood Gas Analyzers.

This recall has been terminated (originally issued August 2, 2006).

Company
Radiometer America Inc
Recall Initiated
June 17, 2005
Posted
August 2, 2006
Terminated
September 14, 2011
Recall Number
Z-1319-06
Quantity
13 boxes/4 membranes per box
Firm Location
Westlake, OH
Official Source
View on FDA website ↗

Reason for Recall

The recalled membranes cause the status value of the pH and electrolyte electrodes to be offset causing readings outside of the acceptable range.

Distribution

Nationwide, including TX, OK, MI, CO, NY, and Canada.

Lot / Code Info

Lot Code: R0308, Expiration Date: 4/07; Lot Code: R0309, Expiration Date: 5/07; and Lot Code: R0310, Expiration Date: 5/07.

Root Cause

Other

Action Taken

The recalling firm notified their distributor/customers of this recall, via telephone on 6/17/2005, at which time, the customers were asked to destroy any and all suspect reference membranes found on-site and at their subaccount customer locations.