D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer.

This recall has been terminated (originally issued August 1, 2006).

Company
Radiometer America Inc
Recall Initiated
February 1, 2005
Posted
August 1, 2006
Terminated
November 18, 2011
Recall Number
Z-1314-06
Quantity
85/4 unit boxes
Firm Location
Westlake, OH
Official Source
View on FDA website ↗

Reason for Recall

The CI membrane units generate too low values for status and sensitivity during calibration.

Distribution

Nationwide.

Lot / Code Info

Lot Code: R0053.

Root Cause

Other

Action Taken

The recalling firm sent a letter, dated 2/05, to all of their consignees.The letter instructed the customers to discontinue use of the product and discard any remaining stock which they had on hand. The letter was accompanied by replacement chloride membranes to replace the customer''s portion of the suspect lot.