ABL 5 Capillary Adaptor (a connection that fits the capillary tube on to the aspirator on the ABL 5 & BPH 5 Analyzers), in vitro diagnostic accessory, Part Number: 906-018.

This recall has been terminated (originally issued August 3, 2006).

Company
Radiometer America Inc
Recall Initiated
June 9, 2005
Posted
August 3, 2006
Terminated
February 24, 2012
Recall Number
Z-1320-06
Quantity
155 bags
Firm Location
Westlake, OH
Official Source
View on FDA website ↗

Reason for Recall

The dimensions and shape of the analyzer adaptor do not fit the analyzer tubes or probe, causing air to be aspirated with the blood sample and causing a deviation in the pO2 sample results.

Distribution

Nationwide, including KS, OH, MD, FL, TX, CA, KY, MA, VA, WA, IL, AR and ME, and Canada.

Lot / Code Info

Part #906-018  lot 041643 to lot 045124

Root Cause

Other

Action Taken

The firm issued a letter to their customers dated 6/9/2005. The letter reports the defect with the ABL5 Capillary Adapters and informs the customers that Radiometer reps will be replacing all of the customers'' ABL5 adaptors ASAP. The letter is accompanied by a single bag of new adaptors for the customers'' ''immediate needs'' and also provides a faxed response form which the customer can use to request additional adapters as required. In addition, should the customer choose to use the old adapters in their possession, the letter stresses and reiterates the ''Precautions and Warnings'' applicable for use of the ABL5 should air bubbles be observed in the measuring chamber by the user/operator.