Potassium Electrode Tip; Part No.s: 668281 Tip Assembly- Synchron Family 669117 Packaged- Synchron Family 441104 Electrode Assembly- EL-ISE 445604 Electrode Packaged- EL-ISE

This recall has been terminated (originally issued June 28, 2005).

Company
Beckman Coulter Inc.
Recall Initiated
May 20, 2005
Posted
June 28, 2005
Terminated
November 21, 2005
Recall Number
Z-0937-05
Quantity
2686
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

Does not meet performance criteria and exhibits upward trend of result recoveries during calibration period.

Distribution

United States and Canada

Lot / Code Info

Multiple lots are involved.    |SYNCRHON: | | |  |Lot 446 ¿ Manufactured 20 NOV 2004* |231 |N/A** |  |Lot 450 ¿ Manufactured 12 DEC 2004 |880 | |  |Lot 501 ¿ Manufactured 03 JAN 2005 |30 | |  |Lot 504 ¿ Manufactured 24 JAN 2005 |698 | |  |Lot 507 ¿ Manufactured 21 FEB 2005 |400 | |  |Lot 508 ¿ Manufactured 28 FEB 2005 |130 | |  |Lot 512 ¿ Manufactured 28 MAR 2005 |1643 | |  |Lot 517 ¿ Manufactured 02 MAY 2005 |2020 | |  | | | |  |EL-ISE: | | |  |Lot 0446 ¿ Manufactured 20 NOV 2004* |20 |N/A** |  |Lot 0504 ¿ Manufactured 24 JAN 2005 |30 | |  |Lot 0508 ¿ Manufactured 28 FEB 2005 |30 | |  |---------------------------------------------+---------------+---------------------|

Root Cause

Other

Action Taken

Firm sent letters by US Mail requesting that users discard thier probes.