Stat Profile Critical Care Xpress CCX (R) Analyzer Catalog Number: 37562 R= Refurbished
- Company
- Nova Biomedical Corporation
- Recall Initiated
- March 30, 2004
- Posted
- July 20, 2004
- Terminated
- June 17, 2004
- Recall Number
- Z-0828-04
- Quantity
- 3 units
- Firm Location
- Waltham, MA
Reason for Recall
Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems
Distribution
CA, IL, LA, MA ,NJ, NY, OH, PA, TX International: Italy ,England, Japan,Turkey., U.A.E., Greece, Thailand, Switzerland, Finland, Saudi Arabia, Italy, Mexico, Trinidad, Ireland, France, Cyprus, Czech Republic, ,Germany
Lot / Code Info
Serial Numbers: Inst. S/N Y01902211 Y01203021 Y01B02101
Root Cause
Other
Action Taken
Nova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.