ABG Replacement Pump Assembly Catalog Number: 37092

Company
Nova Biomedical Corporation
Recall Initiated
March 30, 2004
Posted
July 20, 2004
Terminated
June 17, 2004
Recall Number
Z-0830-04
Quantity
5 units
Firm Location
Waltham, MA

Reason for Recall

Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems

Distribution

CA, IL, LA, MA ,NJ, NY, OH, PA, TX International: Italy ,England, Japan,Turkey., U.A.E., Greece, Thailand, Switzerland, Finland, Saudi Arabia, Italy, Mexico, Trinidad, Ireland, France, Cyprus, Czech Republic, ,Germany

Lot / Code Info

Devices shipped between March 30-31, 2004

Root Cause

Other

Action Taken

Nova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.