VersaRad/Traumex General Radiographic System.

This recall has been terminated (originally issued July 20, 2004).

Company
Fischer Imaging Corporation
Recall Initiated
April 23, 2004
Posted
July 20, 2004
Terminated
March 13, 2012
Recall Number
Z-0974-04
Quantity
138 units
Firm Location
Denver, CO
Official Source
View on FDA website ↗

Reason for Recall

If two of the four T-nuts that connect the transverse rail to the ceiling rail fail, the 'U' arm and transverse rails could fall and cause serious injury.

Distribution

Nationwide, including VA facilities in MI, MN and VT. No military consignees. Foreign distribution to Australia, Canada, Denmark, Italy, Korea, Netherlands, Norway, Sweden, United Kingdom.

Lot / Code Info

All units.

Root Cause

Other

Action Taken

Consignees were notified by letter on 4/23/2004.