VersaRad/Traumex General Radiographic System.
This recall has been terminated (originally issued July 20, 2004).
- Company
- Fischer Imaging Corporation
- Recall Initiated
- April 23, 2004
- Posted
- July 20, 2004
- Terminated
- March 13, 2012
- Recall Number
- Z-0974-04
- Quantity
- 138 units
- Firm Location
- Denver, CO
- Official Source
- View on FDA website ↗
Reason for Recall
If two of the four T-nuts that connect the transverse rail to the ceiling rail fail, the 'U' arm and transverse rails could fall and cause serious injury.
Distribution
Nationwide, including VA facilities in MI, MN and VT. No military consignees. Foreign distribution to Australia, Canada, Denmark, Italy, Korea, Netherlands, Norway, Sweden, United Kingdom.
Lot / Code Info
All units.
Root Cause
Other
Action Taken
Consignees were notified by letter on 4/23/2004.