8F Smart Port CT, Catalog number: CT80STPD and CT80STPD-NF; AngioDynamics, INC. The Smart Port CT Implantable Port is a Single Titanium Port System with Attachable 8.0 Fr. Polyurethane Catheters and 8Fr. Introducer Kit.

This recall has been terminated (originally issued October 13, 2011).

Company
Angiodynamics, Inc
Recall Initiated
August 1, 2011
Posted
October 13, 2011
Terminated
March 7, 2012
Recall Number
Z-0056-2012
Quantity
1095 total units
Firm Location
Manchester, GA
Official Source
View on FDA website ↗

Reason for Recall

The product may contain an incorrect size catheter and introducer.

Distribution

(USA) Nationwide distribution including the states of CA, FL, GA, IA, IL, IN, KY, MD, MI, MO, NC, NE, NJ, NY, OH, OK, OR, TN, UT, WA and WI.

Lot / Code Info

Catalog number: CT80STPD and CT80STPD-NF

Root Cause

Device Design

Action Taken

AngioDynamics, Inc contacted affected customers initially via telephone and a follow up "Notification of Medical Device Correction" letter was sent dated August 1, 2011. The letter informed them of the problem and provided instructions. Customers were instructed to utilize the 8F Introducer Sheath provided within the 8F Smart Port CT Implantable Port during port implantation. The 8F Valved Sheath (V18) may continue to be utilized during the placement of the 7.5F Smart Port Implantable Port (for example: CT75STSD; CT75STSD-NF). For question on this recall contact AngioDynamics, Inc at (800) 772-6446.