VARIANT nbs Sickle Program Resin Update CD-ROM; Model Number:250-3020, included in 250-3000; Distributed and Manufactured by Bio-Rad Laboratories, Inc. The VARIANT nbs Sickle Cell Program is intended as a qualitative screen for the presence of hemoglobins F, A, S, D, C and E in eluates of neonatal blood collected on filter paper by high-performance liquid chromatography(HPLC). The VARIANT nbs Sickle Cell Program is intended for Professional use only.

Company
Bio-Rad Laboratories Inc
Recall Initiated
April 14, 2010
Posted
April 14, 2011
Terminated
August 8, 2011
Recall Number
Z-1984-2011
Quantity
658 units
Firm Location
Hercules, CA

Reason for Recall

The affected resin Update CD-ROMs cause all user-defined settings in the Setup/Test/Data Setup sub screen and Set/Test/Pattern Setting sub screen to be overwritten when the Update Kit procedure is performed.

Distribution

Worldwide Distribution -- USA , Brazil, Canada, Caribbean, France, Italy, and United Kingdom.

Lot / Code Info

Resin Update CD-ROM, 250-3020 lot # AA 00092 goes with Reorder Pack Lot No. 50392618 Resin Update CD-ROM, 250-3020 lot # BA 00092 goes with Reorder Pack Lot No 50300210.  Resin Update CD-ROM, 250-3020 lot # CA 00092  Reorder Pack Lot No 50300645 & 50300650. Expiration date: 05/31/2011, 06/30/2011, 08/31/2011 PIN No. 250-3000;

Root Cause

Software design (manufacturing process)

Action Taken

Bio-Rad Laboratories sent out the Urgent Medical Device Correction notice on 4/14/2011, by Federal Express. An Update was sent on May 13th, advising customers of a minor discrepancy in the parameters included on the affected Resin Update CD-ROMs. Customers were instructed to examine your inventory for the affected reorder pack lots. Please complete and return the Medical Device Correction Response Form Questions regarding this recall please call (510) 724-6114.