Bio-Rad Variant II Turbo Link Hemoglobin Testing System, including Hemoglobin A, Program Reorder Pack 1600 Tests, Program Buffer A, Buffer B, Cartridge Set, Analytical Cartridge, Guard Cartridge, Calibrator, Accessory Box, Instruction Manual and Quick Guide Intended use: Measurement of Hemoglobin A1c in human whole blood.

Company
Bio-Rad Laboratories Inc
Recall Initiated
February 19, 2010
Posted
February 15, 2011
Terminated
March 3, 2011
Recall Number
Z-1259-2011
Quantity
27
Firm Location
Hercules, CA

Reason for Recall

The results from the blood device had the occurrence of some ramping baselines on their chromatograms. The ramping baseline can affect quantitation of the blood results.

Distribution

Worldwide distribution: USA including states of: OH, MN, SC and VA; and country including: Japan.

Lot / Code Info

Catalog No. 270-2716 Lot No. 70291030, expiration date 2/28/2010 Lot No. 70392119, expiration date 3/31/2010 Lot No. 70292119R, expiration date 3/31/2010

Root Cause

Nonconforming Material/Component

Action Taken

The firm, Bio-Rad Laboratories, Inc., sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated February 5, 2010, to all customers and letter dated February 16, 2010 to subsidiaries via Fed-Ex or by fax. The letter described the product, problem and action to be taken by the customers. The customers were instructed to review there chromatograms form the lots listed and compare them to what documented; follow the noted instructions; destroy reorder packs from the affected lots if a ramping baseline issue is observed, and complete and return the attached CUSTOMER MEDICAL DEVICE CORRECTION RESPONSE FORM via fax to (510) 741-3954, attention: Bio-Rad CSD Regulatory Affairs Department. If you have any questions, please contact your regional Bio-Rad Office or (510)724-7000.