Medtronic Sprint Quattro Secure S, model 6935. Steroid eluting, tripolar, screw-in, ventricular lead with RV defibrillation coil electrode. Contents Sterile. Manufactured in: Villalba, Puerto Rico, USA. Medtronic, Inc., Minneapolis, MN 55432 U.S.A.

This recall has been terminated (originally issued November 26, 2010).

Company
Medtronic Inc. Cardiac Rhythm Disease Managment
Recall Initiated
October 25, 2010
Posted
November 26, 2010
Terminated
February 4, 2012
Recall Number
Z-0474-2011
Quantity
154,000 total
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Over--retraction of the helix during initial implant or subsequent repositioning may result in the inability to extend the helix. This does not impact acute or chronic performance of successfully implanted leads.

Distribution

Worldwide Distribution: Throughout the US, including Guam, Puerto Rico, and US Virgin Islands, and to the countries of: Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bangladesh, Barbados, Belgium, Bolivia, Bosnia And Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, El Salvador, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saint Lucia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Viet Nam.

Lot / Code Info

all devices are affected, however, it is an issue that can present itself only at the time of the surgical procedure (acute issue) and it doesn't affect teh chronic performance of a successfully implanted lead.

Root Cause

Device Design

Action Taken

Medtronic issued an Important Medical Device Correction letter, dated October 2010, to doctors. The letter contains a Performance Note. The letter describes the issue and the performance note provides information on how to mitigate the issue through replacement with a new lead. Customers can contact Medtronic Technical Services at 800 723-4636