Huber Plus Non-Coring Needle Safety Infusion Set, 19g, 1 inch, REF 011901, Sterile EO, Assembled In Mexico, Bard Access Systems, Inc., SLC, UT 84116. Safety IV administration set used to access surgically implanted vascular ports.

This recall has been terminated (originally issued February 26, 2010).

Company
Bard Access Systems
Recall Initiated
January 14, 2010
Posted
February 26, 2010
Terminated
May 14, 2010
Recall Number
Z-0862-2010
Quantity
775 units
Firm Location
West Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

After activation of the safety mechanism, the tip of the needle protrudes from the safety sheath creating the potential for needlestick injury.

Distribution

Worldwide Distribution -- United States (VA, KY, MA, NJ and TN) and Belgium.

Lot / Code Info

Lot Number: RETJ0656.

Root Cause

Component change control

Action Taken

Bard Access Systems, Inc. issued and "Urgent: Huber Plus Non-Coring Needle Safety Infusion Set Recall" notification beginning January 2010. Users were instructed to quarantine all affected product and to notify any of their customers that had received the product. Quarantined product is to be returned to the firm. For further information, contact Bard Access Systems, Inc. at 1-800-290-1689.