RecallDepth

PrecisePLAN 2.10 Treatment Planning System

This recall has been terminated (originally issued April 16, 2010).

Company
Elekta, Inc.
Recall Initiated
March 15, 2006
Posted
April 16, 2010
Terminated
December 30, 2010
Recall Number
Z-1406-2010
Quantity
40 units
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

Error in dose rates which were calculated at exactly twice the expected values.

Distribution

Worldwide: IL, NY, PA, TN, Canada, China, Czech Republic, Germany, India, Israel, Italy, Mexico, Poland, Russia, Spain, Switzerland

Lot / Code Info

1762685762, 1762688017, 1762688506, 1726288507, 1762676911, 1762688175, 1762689384, 1762688019, 1762688022, 1762688876, 1762689442, 1762688177, 1762690813, 1762686657, 1762689588, 1762689386, 1762684357, 1762686833, 1762684370, 1762678238, 1762689519, 1762684359, 1762690814, 1762688514, 1762578223, 1762678239, 1762684360, 1762684361, 1762684362, 1762684367, 1762683737, 1762682414, 1762688081, 1762686584, 1762686843, 1762686832, 1762688024, 1762688080, 1762686836, 1762684358

Root Cause

Software design

Action Taken

As described in a prior related submission, Phase I - "Important Notice A285, Scrolling Selected Frame Dose Calculator Error" dated March 15, 2006 was distributed to affected customers to inform them of the potential and what steps should be taken. Phase 2 -- Software release PrecisePLAN 2.11 was released July 11, 2006 to correct this problem. All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. Phase 3 -- All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. The company is directing questions to the local Electa representative or to their offices in the U.K. at telephone 44 (0)-1293-654200.

More recalls by Elekta, Inc.

View all recalls by this company →