RecallDepth

Life shield Latex-Free Primary I.V. Set, yellow key slide clamp, 120 inch, convertible pin, with back check valve, 3 CALV ports, 0.2 micron filter, piggyback with OPTION-LOK, 15 drops/mL; a sterile Rx single use device; Homspera Inc, Lake Forest, IL USA. Made in Costa Rica; list 20793-48.

This recall has been terminated (originally issued April 15, 2010).

Company
Hospira Inc.
Recall Initiated
November 24, 2009
Posted
April 15, 2010
Terminated
January 29, 2012
Recall Number
Z-1382-2010
Quantity
7,104 sets
Firm Location
Lake Forest, IL
Official Source
View on FDA website ↗

Reason for Recall

Reports of difficulty in adjusting the flow rate, which could result in flow rate variances.

Distribution

Worldwide Distribution -- U.S., Puerto Rico, Dominican Republic, Bahamas, Virgin Islands, Argentina, Australia, Bahrain, Belgium, Canada, China, Denmark, Ecuador, EI Salvador, Finland, France, Germany, Greece, Guatemala, Hong Kong, Ireland, Italy, Jordan, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uruguay, and United Arab Emirates.

Lot / Code Info

Lot numbers: 610155H, 630265H, 641015H.

Root Cause

Component design/selection

Action Taken

Hospira sent the Urgent Device Recall letters dated 11/25/09 to their direct accounts of the affected products and requested these consignees to notify their customers of the recall. The accounts were informed that Hospira ceased manufacturing the I.V. gravity and pump administration sets with the CAIR clamp manufactured by Borla S.P.A. of Italy in March 2008, due to customer reports related to a difference in the zone of control used to adjust the flow compared to the original CAIR clamp design. To avoid the possibility of flow rate variances, Hospira is recalling all remaining sets with the Borla designed CAIR clamp. The accounts were requested to quarantine all inventory of the affected lots, complete the enclosed reply form indicating the amount of product being returned, fax a copy of the completed form to Stericycle at 1-877-377-5114, and include a copy with the returned goods to Stericycle at 2670 Executive Drive, Suite A, Indianapolis, IN 46241.

More recalls by Hospira Inc.

View all recalls by this company →