Leksell Gamma Knife. Leksell Gamma Knife is a teletherapy device indicated for use in the stereotactic irradiation of intracranial structures

This recall has been terminated (originally issued April 13, 2010).

Company
Elekta, Inc.
Recall Initiated
October 1, 2007
Posted
April 13, 2010
Terminated
December 28, 2010
Recall Number
Z-1236-2010
Quantity
5 units
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

After updating the LGK actuator in spare part 810361, the old sleigh became obsolete due to causing insufficient locking of the helmet in combination with the new actuator.

Distribution

United States (CA, TN, TX and WA) and Taiwan.

Lot / Code Info

LGK 4203, LGK 4309, LGK 4333, LGK 4343 and LGK 4354.

Root Cause

Other

Action Taken

Field Change Order 0000203 "Replacement of obsolete actuator sleigh on Helmet Changer", was sent to all customers that were affected. Direct questions to Elekta, Inc. by calling 1-770-670-2548.