Smiths Medical, Add-On Kids Kit , Blood Sampling System, Product Code: MX4033, Smiths Medical ASD, Inc, Dublin, Ohio 43016

This recall has been terminated (originally issued January 31, 2008).

Company
Smiths Medical Asd Inc
Recall Initiated
November 21, 2007
Posted
January 31, 2008
Terminated
September 30, 2010
Recall Number
Z-0696-2008
Quantity
250 units
Firm Location
Dublin, OH
Official Source
View on FDA website ↗

Reason for Recall

Misbranded; The product codes are labeled incorrectly with a green promotional label for "New split septum sampling site. Only use needle-free access devices .......", however this product code contains a sampling site which is accessed with a shrouded needle, NOT with a needle-free access device

Distribution

Nationwide including the states of CA, DC, IL, IN, MD, MN, NC, NE, SC, and TN and Puerto Rico.

Lot / Code Info

Lot Numbers: 1189082, 1231444, 1178100, and 1225588.

Root Cause

Packaging process control

Action Taken

Smiths Medical mailed certified letters (Urgent Product Advisory Notifications) dated November 19, 2007, to all customers that received the affected lot. Firms are instructed to make the field correction (extra label removal) or mail the product back to Smiths for the correction.