Liaison 25 OH Vitamin D Assay, Manufacturer part #310900, Lot #114668

Company
Diasorin Inc.
Recall Initiated
February 14, 2005
Posted
March 18, 2005
Terminated
July 22, 2006
Recall Number
Z-0638-05
Quantity
31 kits
Firm Location
Stillwater, MN

Reason for Recall

An internal inspection indicates that Calibrator 2 and Conjugate are switched in their positions in the integral. This will result in Calibrator 1 RLU's similar to that of background RLU, and Calibrator 2 RLU's near expected range. A curve can not be calculated from this data.

Distribution

MA, Spain, Germany & Belgium

Lot / Code Info

manufacturer part # 310900, Lot #114668

Root Cause

Other

Action Taken

DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.