LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E.

Company
Diasorin Inc.
Recall Initiated
February 14, 2005
Posted
March 18, 2005
Terminated
July 22, 2006
Recall Number
Z-0637-05
Quantity
131
Firm Location
Stillwater, MN

Reason for Recall

Results of an internal inspection show Lot 114439E to have calibrator 1 and Calibrator 2 switched in their position in the integral. This results in an inability to calculate a curve, due to the reversal of calibrator values.

Distribution

MI, NJ, Germany, Spain and Belgium

Lot / Code Info

catalog no. 310900, lot 114439E

Root Cause

Other

Action Taken

DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.