Deltec CADD-Prizm PCS II Ambulatory Infusion Pump with Revision E Software

Company
Deltec, Inc
Recall Initiated
June 23, 2003
Posted
July 3, 2003
Terminated
July 2, 2004
Recall Number
Z-0980-03
Quantity
135 pumps
Firm Location
St Paul, MN

Reason for Recall

Pump keyboard entries by the patient could have resulted in the patient having unintended access to programming screens and led to in improper drug dosage.

Distribution

The product was distributed only to two hospitals, one in North Carolina and one in Washington State.

Lot / Code Info

Pump Serial numbers 658825, 658837, 658840, 658842, 658844 - 658865, 658867, 658869 - 658896, 658898 - 658918, 658920 - 658936, 658941, 658943 - 658947, 658949, 662473 - 662477, 662483, 662485, 662488 - 662490, and 662635 - 662659

Root Cause

Other

Action Taken

A recall letter, sent to the consignees on June 23, stated that the firm would telephone the consignees to arrange the correction of their affected pumps by reprogramming with the new software.