Elekta Synergy XVI 4.5 Product Usage: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.

This recall has been terminated (originally issued November 29, 2011).

Company
Elekta, Inc.
Recall Initiated
March 9, 2011
Posted
November 29, 2011
Terminated
August 9, 2012
Recall Number
Z-0305-2012
Quantity
56
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

If "Confirm" or "Unconfirm" settings are clicked during the "Terminated Checking" phase of treatment delivery, Desktop Pro cancels the loaded beam and sets the Beam MUs to zero. Desktop Pro then sends the delivered MUs to MOSAIQ a a zero value.

Distribution

Nationwide Distribution (USA) including the states of: AZ, CA, GA, IL, LA, MD, MA, MI, MN, NJ, NY, NC, OH, PA, TN, VA, WI and Puerto Rico

Lot / Code Info

152019, 152448, 151076, 151954, 152454, 152117, 152452, 152393, 152394, 151942, 152163, 151955, 152176, 152222, 151931, 152388, 151141, 152365, 152303, 152358, 152353, 152359, 152352, 152362, 152305, 152468, 152481, 152469, 151812, 152432, 151260, 151952, 152370, 151396, 152301, 152248, 152425, 151155, 151156, 151157, 151412, 105984, 152307, 152308. 151443, 151545, 151628, 152207, 152271, 152373, 152371, 151080, 152211, 152158, 152392, 152331

Root Cause

Software design

Action Taken

Elekta sent an IMPORTANT NOTICE letter dated March 9, 2011 to affected customers. The notice identified the product, the problem, and the corrective action to be taken by the customer. The notice advise all customers to follow any instructions or recommendations covered in the Notice. For questions call +44(0)1293 654200.