Welch Allyn AED 10 automated external defibrillator and MRL Jumpstart (collectively "AED 10")

This recall has been terminated (originally issued March 9, 2009).

Company
Welch Allyn Protocol, Inc
Recall Initiated
December 15, 2008
Posted
March 9, 2009
Terminated
March 19, 2012
Recall Number
Z-1006-2009
Quantity
10, 735
Firm Location
Beaverton, OR
Official Source
View on FDA website ↗

Reason for Recall

Reliability issues - potential to shock a non shockable rhythm or not shocking a shockable rhythm;

Distribution

Worldwide Distribution.

Lot / Code Info

Software version 2.02 or lower. The ECG analysis/noise issue involves all units manufactured before January 2006 when a design/component change was made.

Root Cause

Component design/selection

Action Taken

Welch Allyn sent a Customer Recall Notice dated February 26, 2009, to all customers (customer list supplied by distributors). Welch Allyn will upgrade all units with software version 2.02 and lower to version 2.06.02 by Proc. AED10 Prod Service Min HW/SW Update Rev. E 820-0178-00.