AxSYM Total B-hCG Reagent List 7A59-21 is composed of one bottle of Anti B-hCG Coated Microparticles, one bottle of Anti- B-hCG Alkaline Phosphatase Conjugate and one bottle of Specimen Diluent. REF 7A59-21. 100 Tests. For In Vitro Diagnostic Use.

This recall has been terminated (originally issued July 12, 2007).

Company
Abbott Diagnostic International, Ltd. 30 Bo Cruce Davila
Recall Initiated
April 10, 2007
Posted
July 12, 2007
Terminated
February 14, 2012
Recall Number
Z-1067-2007
Quantity
11530 kits
Firm Location
Barceloneta, PR
Official Source
View on FDA website ↗

Reason for Recall

An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is magnified potentially yelding false positive results for pregnancy and or Dilution Inaccuracy.

Distribution

Worldwide, including USA, PR, USVI, CANADA, MEXICO, GUATEMALA, COSTA RICA, COLOMBIA, ECUADOR, PERU, CHILE, URUGUAY, ARGENTINA, BRAZIL, DOMINICAN REPUBLIC, EL SALVADOR, GERMANY, JAPAN, SINGAPORE, HONG KONG, TAIWAN, THAILAND, SOUTH KOREA, AUSTRALIA, NEW ZEALAND, HONDURAS, TRINIDAD & TOBAGO, GRENADA, ST. VINCENT, ST. LUCIA, ANTIGUA, BARBADOS, CURACAO, BRITISH VIRGIN ISLANDS, BAHAMAS, PANAMA, and CAYMAN ISLANDS.

Lot / Code Info

46329Q101, 47118Q101, 48063Q101, 48065Q101, 49341Q101, 49559Q101

Root Cause

Other

Action Taken

The 4/10/2007 Product Recall Letter was sent to all customers who received the affected lots.