Access Immunoassay Systems Thyroglobulin Antibody Reagent Kit, Part Number 33890

This recall has been terminated (originally issued February 7, 2007).

Company
Beckman Coulter Inc.
Recall Initiated
October 26, 2006
Posted
February 7, 2007
Terminated
April 30, 2012
Recall Number
Z-0417-2007
Quantity
474 units
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

Low end imprecision affecting both S0 calibrators and patient samples which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay.

Distribution

Nationwide and Canada

Lot / Code Info

518729 (expires 10/31/06); 520617 (expires 12/31/06); 612764 (expires 12/31/06); 613481 (expires 3/31/07); 614541 (expires 5/31/07); 616787 (expires 8/31/07)

Root Cause

Other

Action Taken

A Product Corrective Action (PCA) letter was mailed on Oct 27, 2006 to all Thyroglobulin Antibody customers. They are informed that there has been an increase in customer feedback regarding low end imprecision affecting both S0 calibrators and patient samples, which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay. The letter informs the customer that the firm recommends that they use an alternate method for testing TgAb. (Letter was sent by US mail). Customers are asked to complete and return the enclosed response form within 10 days.