Disposable Tips 1000 l with filter Conductive disposable tips for Tecan Genesis Series * Order Number 612513.1 * Tecan Schweiz AG Seestrae 103, CH-8708 Mannedorf, Switzerland * Made in Germany. Blister 1 x 96 pcs.

This recall has been terminated (originally issued December 23, 2005).

Company
Tecan U S Inc
Recall Initiated
April 11, 2005
Posted
December 23, 2005
Terminated
May 18, 2006
Recall Number
Z-0326-06
Quantity
3000 Tips
Firm Location
Research Triangle Park, NC
Official Source
View on FDA website ↗

Reason for Recall

Double filtering of 1000 l tips. The defect could result in cross contamination of other samples due to possible dripping of sample while moving over other samples.

Distribution

IL and Europe

Lot / Code Info

Part Number 10612523, Lot No. U122098L-1325, Lot No. U122098L-3225 & Lot No. U122098L-2325. The sub-lot numbers (i.e. 1325, 3225 and 2325) are printed on each blister and stamped in blue colour on the shipping container in the upper right corner of the box.

Root Cause

Other

Action Taken

The consignee was notified via fax''d letter on/about 11/08/2005.