Spectrum Patient Monitor, Physiological, Patient with arrhythmia detection or alamrs. Monitor can monitor, display, trend and print a patient''s physiological parameters. The device has a 12.1 inch color display and has a standard configuration of a 3 or 5 lead ECG, Masimo SET SpO2, non-invasive blood pressure, respiration, continuous temperature and IV Drug Calculations. Optional digital displays are provided for invasive blood pressure (up to four) anesthetic agents, O2 and NO2, ST, and CO2.

This recall has been terminated (originally issued November 23, 2005).

Company
Datascope Corp.
Recall Initiated
September 7, 2005
Posted
November 23, 2005
Terminated
June 6, 2007
Recall Number
Z-0176-06
Quantity
2318 monitors
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data.

Distribution

The monitors have been distributed domestically and internationally to hospitals and clinics. There are 10 Passport 2 monitors distributed to Naval Hospital, Camp Pendleton, CA. There are 12 Spectrum monitors distributed to two US Govt facilities: VA Medical Center, Hines, IL and VA Medical Center, Syracuse, NY.

Lot / Code Info

All Spectrum Monitors that have the following Software Versions: A.19, A.20, B.23, B.24, B.25, B.27, C.18, D.07, E.07, F.09, F.16, F.17, F.26.

Root Cause

Other

Action Taken

Datascope sent out a Field Correction notice on 9/7/2005 to all Hospital administrators who have the monitors. This was sent out via certified mail, return receipt requested. A Datascope Service Rep will conduct and document the corrective action during a visit to the customer.