Hamamatsu model L10951-35 Micro Focus x-ray generator. The x-ray generator is incorporated into a cabinet x-ray system that is used for a variety of studies.

This recall has been terminated (originally issued September 13, 2011).

Company
Hamamatsu Corporation
Recall Initiated
July 7, 2011
Posted
September 13, 2011
Terminated
January 29, 2018
Recall Number
Z-3076-2011
Quantity
8
Firm Location
Bridgewater, NJ
Official Source
View on FDA website ↗

Reason for Recall

It was discovered about a potential issue of X-ray leakage with our Model # LI1 0951-35 Micro Focus X-Ray generator.

Distribution

Worldwide Distribution -- USA ( nationwide ) the countries of Aichi, India and Japan.

Lot / Code Info

Model # L10951-35 Serial number SII00002 through SII00009.

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Hamamatsu Corporation sent a letter dated July 19, 2011, to the manufacturer and their subcontractor informing them of the affected product and what they propose to do. They will repair the units both in the field and those returned to our plant by installing additional lead shielding in the units. The lead shielding will be fixed to the unit by soldering the addition to the existing lead shield (see attachment 2). We expect to complete the repair of all units in the field by October 1, 2011. All new units to be shipped from our plant will be repaired before shipment. For further questions, please call (908) 231-0960 X 2300