HomeChoice and Yume Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C4471, 5C4471R, 5C4474, 5C4474R, T5C4441, T5C4441R

This recall has been terminated (originally issued July 20, 2004).

Company
Baxter Healthcare Renal Div
Recall Initiated
May 24, 2004
Posted
July 20, 2004
Terminated
December 11, 2007
Recall Number
Z-1012-04
Quantity
9,026 units
Firm Location
Mc Gaw Park, IL
Official Source
View on FDA website ↗

Reason for Recall

A software defect could result in a patient experiencing an overfill with software versions 8.5 and higher. The possibility of an overfill can occur when the Initial Drain Alarm Volume is set to OFF and a change to the volume is made at the Verify I-DRAIN:XXXXML prompt.

Distribution

Nationwide, Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Denmark, Ecuador, France, Germany, Guatemala, Hong Kong, India, Ireland, Italy, Korea, Malaysia, Mexico, Nicaragua, the Netherlands, Peru, the Philippines, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uruguay, Venezuela and the United Kingdom.

Lot / Code Info

All HomeChoice Systems with software versions 8.5 and higher.

Root Cause

Other

Action Taken

Urgent Device Correction letters dated 5/24/04 were mailed to the peritoneal dialysis nurse of facilities identified as having patients who use the HomeChoice machines. The letters informed them of the potential overfill problem and requested that they review the patient list provided with the letter and verify that it was complete. The nurses were requested to communicate the information to their home patients, and visit all of their patients who use the device in the low I-Drain Alarm Volume mode to make an interim correction, permanently changing the I-Drain Alarm Volume from OFF to 0ML in the Make Adjustments menu. Permanent software corrections will be installed when the revised software becomes available. Follow-up letters dated 1/12/05 were sent to the Home Care Coordinators, providing them with an addendum to the HomeChoice Trainer's guide, repeating the information from the 5/24/04 letter and providing additional information to safely disconnect from the HomeChoice system during therapy. The letters also included copies of the 1/14/05 letter sent directly to the HomeChoice home patients, as well as addenda to the Patient Help Cards and Patient At-Home Guide.