North Star Imaging Inc. (NSI) Model X-7000 industrial cabinet x-ray system. Industrial x-ray imaging of product hardware

Company
North Star Imaging Inc
Recall Initiated
August 23, 2017
Recall Number
Z-0063-2018
Quantity
US - 3 units
Firm Location
ROGERS, MN

Reason for Recall

It was discovered a low-level radiation leak that was found on the NSI X-7000 system.

Distribution

US Distribution to the states of : PA, OH and KS.

Lot / Code Info

Model X-7000

Root Cause

Radiation Control for Health and Safety Act

Action Taken

North Star Imaging will bring defect into compliance by the following : The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 3 units will consist of the following: 1. All three customers have been notified that the affected products will be repaired free of charge. 2 .The area where the emission rate could be exceeded will have shielding added until emissions are adequately attenuated under the conditions described above. 3. The repaired systems will be completely surveyed to verify compliance with the cabinet x- ray system performance standards emission limit. 4.All repairs will be completed before February 1, 2018. For further questions, please call (800) 635-8392.