The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrically-powered components intended for imaging and analysis of soft-tissue and vasculature using combined optoacoustic (photoacoustic) and ultrasound (US) imaging.

This recall is currently active, issued February 7, 2025. It was issued by Ithera Medical Gmbh Zielstattstr. 13 Munich Germany.

Company
Ithera Medical Gmbh Zielstattstr. 13 Munich Germany
Recall Initiated
January 8, 2025
Posted
February 7, 2025
Recall Number
Z-1100-2025
Quantity
23 units (6 units US)
Official Source
View on FDA website ↗

Reason for Recall

The MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD).

Distribution

Worldwide distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Germany, Italy, Netherlands, Portugal, Switzerland, and United Kingdom.

Lot / Code Info

Model numbers MSOT Acuity Echo (Research System) MSOT Acuity Echo CE MSOT Acuity Echo CE2 UDI's: (01)04262380070004 (01)04262380070004(21)22102 (01)04262380070004(21)22203 (01)04262380070004(21)21901 (01)04262380070004(21)22101 (01)04262380070004(21)22104 (01)04262380070004(21)22105 (01)04262380070004(21)22106 (01)04262380070004(21)22107 (01)04262380070004(21)22201 (01)04262380070004(21)22202 (01)04262380070004(21)22204 (01)04262380070004(21)22205 (01)04262380070004(21)22206 (01)04262380070004(21)22301 (01)04262380070004(21)22401

Root Cause

Radiation Control for Health and Safety Act

Action Taken

The firm sent an "Urgent Field Safety Notice" letter dated 12/19/2024 to its consignees. FSN 3746 has been distributed to all customers to correct a wrong value in the user manual for the NOHD. The letter described the product, problem and actions to be taken. The consignees were instructed to do the following: Take note of amendment/reinforcement of Instructions For Use (IFU) 1. Confirm the receipt of this FSN 2. Forward this information and the updated IFU to the person responsible for laser safety at your location (such as Laser Safety Officer) 3. The person responsible for laser safety at your location performs a new assessment of the laser safety consideration and to identify any additionally required work safety related measures 4. Training of all users of the MSOT Acuity Echo device on any additionally required work safety related measures, if applicable. The new calculation of the NOHD resulting in a higher value due to the use of a different calculation method. The user manual has been updated accordingly, and users have been informed. iThera will provide an updated user manual (IM_2.03_EN_III) with the corrected value for NOHD. If you have any questions, call 203-997-5110 US, +49-80-700749; EXT21 OUS or email: josef.konardl@ithera-medical.com.