MAVERICK Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Company
Repligen Corporation
Recall Initiated
December 10, 2025
Posted
January 13, 2026
Recall Number
Z-1024-2026
Quantity
33
Firm Location
Marlborough, MA

Reason for Recall

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Distribution

U.S.

Lot / Code Info

PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Repligen will repair the connections in each potentially affected MAVERICK device. In the meantime, all customers have been instructed to refrain from using the device.