IBA Proton Therapy System - PROTEUS 235

Company
Ion Beam Applications S.A. Chemin Du Cyclotron 3 Ottignies-Louvain-La-Neuve Belgium
Recall Initiated
February 3, 2026
Posted
March 6, 2026
Recall Number
Z-1497-2026
Quantity
4 units (1 US, 3 OUS)

Reason for Recall

It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.

Distribution

Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.

Lot / Code Info

PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)

Root Cause

Software design

Action Taken

On February 3, 2026 Urgent Medical Device Correction letters were sent to customers. Letter provides awareness of the Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. On-site modification of the devices will be performed by IBA with a TSS Configuration update. The update will be deployed at the impacted sites by June 2026.