Medline medical procedure convenience kits labeled as: CIRCUMCISION PACK, Kit SKU DYNJ16826O

Recall Initiated
December 12, 2025
Posted
January 28, 2026
Recall Number
Z-1198-2026
Quantity
1,928 total
Firm Location
Northfield, IL

Reason for Recall

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.

Distribution

US Nationwide distribution.

Lot / Code Info

Medline Kit SKU DYNJ16826O: UDI/DI 10195327412067 (EA) 40195327412068 (CS), Lot Number 25LMA655; Medline Kit SKU DYNJ16826O: UDI/DI 10195327412067 (EA) 40195327412068 (CS), Lot Number 25KMC911.

Root Cause

Nonconforming Material/Component

Action Taken

Medline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-25-251-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.