Philips Incisive CT

Company
Philips Healthcare (Suzhou) Co., Ltd. Suzhou Industrial Park No.
Recall Initiated
December 3, 2025
Posted
January 8, 2026
Recall Number
Z-1003-2026
Quantity
35
Firm Location
Pingjiang Suzhou China

Reason for Recall

Potential for incomplete scan due to unstable connection inside of floating sensor.

Distribution

U.S. and U.S. territories

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide the actions that should be taken in order to prevent risks to patients. Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit their site and install a hardware solution that will change the method of connection for the rear handle floating sensor. Customers who need additional information or support concerning this issue should be directed to their local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center). Philips reference: 2025-PD-CTAMI-109 or FCO72800843.