RecallDepth

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.

This recall is currently active, issued October 10, 2025.

Recall Initiated
September 11, 2025
Posted
October 10, 2025
Recall Number
Z-0132-2026
Quantity
174,013 units
Firm Location
Nagog Park Acton, MA
Official Source
View on FDA website ↗

Reason for Recall

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Distribution

US Nationwide distribution.

Lot / Code Info

Omnipod 5 App Software. UDI: 10385083000459. All software versions 2.X.X

Root Cause

Software design

Action Taken

Insulet notified both healthcare professionals and users on about 09/11/2025 via URGENT: Medical Device Correction letter. Instructions included that an update to the application will be pushed to all devices and should be implemented by users. Users were also requested to confirm receipt and acknowledgement of the notification but selecting the "acknowledgement button" in the email.