RecallDepth

MOSAIQ Oncology Information System Software Builds 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 Software Version and UDI Codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500026(10)3.1.3.2 3.1.3.3 (01)07340201500026(10)3.1.3.3 3.1.3.4 (01)07340201500026(10)3.1.3.4 3.2.1.0 (01)07340201500071(10)3.2.1.0 3.2.1.1 (01)07340201500071(10)3.2.1.1 3.2.1.2 (01)07340201500071(10)3.2.1.2 3.2.1.3 (01)07340201500071(10)3.2.1.3 3.2.1.4 (01)07340201500071(10)3.2.1.4 3.2.2.0 (01)07340201500071(10)3.2.2.0 3.2.2.1 (01)07340201500071(10)3.2.2.1 3.2.3.0 (01)07340201500071(10)3.2.3.0 MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.

This recall is currently active, issued November 14, 2025. It was issued by Elekta Solutions Ab.

Company
Elekta Solutions Ab
Recall Initiated
September 17, 2025
Posted
November 14, 2025
Recall Number
Z-0503-2026
Quantity
4 systems that contain the affected software.
Firm Location
Stockholm Sweden
Official Source
View on FDA website ↗

Reason for Recall

When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule.

Distribution

Worldwide - U.S. Nationwide distribution in the state of NM and the countries of Canada, Malta, and South Africa.

Lot / Code Info

Software Versions: 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 UDI codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500026(10)3.1.3.2 3.1.3.3 (01)07340201500026(10)3.1.3.3 3.1.3.4 (01)07340201500026(10)3.1.3.4 3.2.1.0 (01)07340201500071(10)3.2.1.0 3.2.1.1 (01)07340201500071(10)3.2.1.1 3.2.1.2 (01)07340201500071(10)3.2.1.2 3.2.1.3 (01)07340201500071(10)3.2.1.3 3.2.1.4 (01)07340201500071(10)3.2.1.4 3.2.2.0 (01)07340201500071(10)3.2.2.0 3.2.2.1 (01)07340201500071(10)3.2.2.1 3.2.3.0 (01)07340201500071(10)3.2.3.0

Root Cause

Software design

Action Taken

On 09/17/2025, the firm emailed an "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" to customers informing them when appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule. Customers are instructed to: When appending cycles for care plans that contain additional medication orders with a wave schedule, review and adjust the appended care plan cycles to reflect the clinical intent. Alternatively, wave orders can be removed from the care plan folder. Additional information can be found in: Customers with v3.1.3 - section 3.4.1 - page 35-36 in doc E062133 / 01 MOSAIQ 3.1.3.0 Instructions for Use Customers with v3.2.1 - section 3.4.1 - page 35-36 in doc E065117 / 03 MOSAIQ 3.2.1.0 Instructions for Use Customers with v3.2.2 - section 3.4.1 - page 35-36 in doc E067652 / 04 MOSAIQ 3.2.2.0 Instructions for Use Customers with v3.2.3.0 - section 3.4.1 - page 35-36 in doc E071967 / 01 MOSAIQ 3.2.3.0 Instructions for Use For questions - contact Quality Assurance at +46 70-962-0862 or email mats.premfors@elekta.com