Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 530 MD, Rx Only

This recall is currently active, issued October 21, 2025. It was issued by Uih Technologies Llc.

Company
Uih Technologies Llc
Recall Initiated
September 12, 2025
Posted
October 21, 2025
Recall Number
Z-0217-2026
Quantity
8 units
Firm Location
Houston, TX
Official Source
View on FDA website ↗

Reason for Recall

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Distribution

US: OUS:

Lot / Code Info

Serial Numbers: ; UDI: (01)06971576831012/ Serial Numbers: 300162 300163 300174 300176 304002 304003 304004 304006

Root Cause

Other

Action Taken

On September 12, 2025, UIH Technologies LLC, issued a recall notification to affected consignees. UIH asked consignees to take the following actions: 1. Ensure patient positioning clearance: When positioning the patient, confirm that arms and legs and clear of all table gaps and moving parts. 2. Avoid use of third-party slickers: We advise against the use of any non-UIH-approved slickers. 3. Service Contact: For technical assistance, contact your United Imaging Service engineer or reach out to our 24/7 Customer Contact Center, Toll-Free: (855) 221-1552/Direct: (832) 699-2799 4.Report adverse events: if any adverse events or quality issue occur, please report them to the FDA via the MedWatch program. 5. Please share this information with all relevant personnel at your facility, including technologist, service staff, and clinical teams.