Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

This recall is currently active, issued July 29, 2025. It was issued by Capnia Inc.

Company
Capnia Inc
Recall Initiated
April 21, 2025
Posted
July 29, 2025
Recall Number
Z-2198-2025
Quantity
2160
Firm Location
Foster City, CA
Official Source
View on FDA website ↗

Reason for Recall

Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.

Distribution

US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.

Lot / Code Info

Monitor UDI-DI: 00858481006017. Lot (C20206): 231220

Root Cause

Device Design

Action Taken

Starting on 4/21/2025 Product Advisory and Voluntary Field Action emails were sent to customers who were told affected devices would be replaced. Customers with questions could contact the firm at 904-588-2986