Akreos, SKU: AO60P0300. Akreos intraocular lenses

Company
Bausch & Lomb Surgical, Inc.
Recall Initiated
June 6, 2025
Posted
August 26, 2025
Recall Number
Z-2451-2025
Quantity
2
Firm Location
Clearwater, FL

Reason for Recall

An off-power error was identified, which resulted in a diopter of 16, rather than the intended 3 diopter. Implantation of affected intraocular lens could result in functional visual impairment or reduction in visual acuity which could result in the need for optical or surgical correction.

Distribution

US: AZ, CA

Lot / Code Info

UDI-DI: 10757770501783, Lot: 1Q29212, Serial Numbers: 1Q29212006 and 1Q29212010

Root Cause

Process control

Action Taken

6/6/2025, recall notices were emailed to customers who were asked to do the following: 1) Check your inventory, quarantine, and do not use the IOL from this impacted lot. 2) Provide this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred (as appropriate). Also, please transfer this notice to other organizations on which this action has an impact (as appropriate). Please maintain awareness of this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. 3) Complete and return the acknowledgement form via email to BLSurgCustSvc@Bausch.com For questions regarding this notice, please call firm at 1-800-338-2020, Option 9.