Atrium Oasis Drain Single W / AC, Part No. 3600-100

This recall is currently active, issued December 28, 2023. It was issued by Atrium Medical Corporation.

Company
Atrium Medical Corporation
Recall Initiated
November 1, 2023
Posted
December 28, 2023
Recall Number
Z-0621-2024
Quantity
835 cases total (6 units/case)
Firm Location
Merrimack, NH
Official Source
View on FDA website ↗

Reason for Recall

A sterilization nonconformance prematurely aged the product by 48.3 days, resulting in the expiration date on the label not being accurate. This nonconforming product was inadvertently distributed to customers. Increased risk to patients and healthcare providers using the product after the shortened expiration date of November 17, 2025.

Distribution

Domestic distribution nationwide. No foreign distribution.

Lot / Code Info

UDI-DI 00650862110012 Lots 492065 and 492075

Root Cause

Nonconforming Material/Component

Action Taken

On November 1, 2023, the firm issued URGENT MEDICAL DEVICE - Removal letters to affected customers. Customers were instructed to remove any product from the affected lots from areas of use. Customers with affected product are entitled to a replacement at no cost to their facility. Customers will receive replacements upon their acknowledgement that they have affected product for return. Distributors should forward recall documentation to affected customers. Customers with questions may contact their Atrium/Getinge representative or call the Atrium/Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, 9AM-6PM Eastern Time Zone.