Achieva 1.5T Conversion, Model No. 781283

This recall is currently active, issued January 25, 2024. It was issued by Philips North America.

Company
Philips North America
Recall Initiated
December 29, 2023
Posted
January 25, 2024
Recall Number
Z-0808-2024
Quantity
7 US; 16 ROW total
Firm Location
Cambridge, MA
Official Source
View on FDA website ↗

Reason for Recall

An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.

Distribution

Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.

Lot / Code Info

Model No. 781283 UDI-DI 00884838009806 Serial Number 39016

Root Cause

Under Investigation by firm

Action Taken

On December 29, 2023, the firm notified affected customers with URGENT Medical Device Correction letters. The firm stated there is no impact to patients or users. There is a potential risk of electrical shock for service personnel. Customers may continue to use MR systems in accordance with the intended use. Circulate this notice to all users of this device so that they are aware of the issue, and post the letter in the technical room until the electrical shock warning label is updated. A Philips representative will contact you to schedule a time for a Field Service Engineer to verify the electrical shock warning label and apply the correct label if applicable. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).