RecallDepth

Solea Models 2.0 and 3.0 Laser Surgical Instrument, Foot Pedal assembly SUB-00434 with foot pedal CUS-00226 Rev. K reworked

This recall is currently active, issued November 17, 2023. It was issued by Convergent Dental.

Company
Convergent Dental
Recall Initiated
September 1, 2023
Posted
November 17, 2023
Recall Number
Z-0324-2024
Quantity
39 units
Firm Location
Needham, MA
Official Source
View on FDA website ↗

Reason for Recall

It has been found that potential unintended laser activation can occur without foot pedal depression.

Distribution

Worldwide - US Nationwide distribution in the states of AK, AL, CA, CT, FL, KY, ME, MI, MO, MT, NC, OH, OR, PA, TN, TX, UT, VA, WA, WI and the country of Canada.

Lot / Code Info

Solea Model 3.0: UDI-DI: 00850634007018; Serial Numbers: 0310516, 0310517, 0310518, 0310519, 0310520, 0310521, 0310522, 0310523, 0310524, 0310525, 0310526, 0310527, 0310528, 0310529, 0310530, 0310531, 0310532, 0310533, 0310534, 0310535, 0310536, 0310537, 0310538, 0310539, 0310540, 0310541, 0310542, 0310543, 0310317, 0310321, 0310322, 0300730; Solea Model 2.0: UDI-DI: 00850634007001; Serial Numbers: 0200822, 0200807, 0200810, 0200625, CDS2-U33W-1B-DE-15, 0200819, 0200401

Root Cause

Under Investigation by firm

Action Taken

An URGENT: MEDICAL DEVICE RECALL notification letter dated 9/1/23 was sent to customers. Please discontinue use of this product until your replacement foot pedal arrives. A replacement foot pedal (part number CUS-00226 Rev Q) will be shipped to the address that we have on file for your office and should arrive no later than September 6, 2023. Your shipment tracking number is [Insert]. A return shipping label will be included in the package. Please send back your original foot pedal for inspection and monitored disposal. Connecting the replacement foot pedal will resolve this issue with no further action required. The part number and sub part number can be found at the bottom of your Solea foot pedal as shown in the attached photo. If the part number, sub-part number, and revision number do not match the recalled product, please call me directly at 617-480-3213. Convergent Dental is committed to product quality as well as patient and user safety. If you have any questions about this recall, please contact our Customer Support line at 508-500-5670. An Acknowledgement and Receipt Form is attached. Please complete the form and return it to Convergent Dental. An amended URGENT: MEDICAL DEVICE RECALL was sent to customers on 9/5/23. Protective eyewear must be worn by everyone who is in the operating environment while the laser is turned on (Armed State). Periodically inspect protective eyewear (laser safety glasses) for pitting and cracking. Please be warned that failure to follow these measures may lead to ocular damage.